Company record
Securitas Healthcare LLC
0 FDA warning letters since 2023 and 2 recalls in the last 12 months.
Get an email the day FDA posts anything new about Securitas Healthcare LLC.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Jan 14 | Class II | Securitas Healthcare LLC Device |
Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3 Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing. FDA notice Z-1032-2026 |
| Jan 14 | Class II | Securitas Healthcare LLC Device |
Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3 Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing. FDA notice Z-1031-2026 |